Periodic safety update report
Manufacturers of class C and class D devices shall prepare a periodic safety update report (‘PSUR’) for each device and where relevant for each category or group of devices summarising the results and conclusions of the analyses of the post-market surveillance data gathered as a result of the post-market surveillance plan referred to in Article 79 together with a rationale and description of any preventive and corrective actions taken. Throughout the lifetime of the device concerned, that PSUR shall set out:
the conclusions of the benefit-risk determination;
the main findings of the PMPF; and
the volume of sales of the device and an estimate of the size and other characteristics of the population using the device and, where practicable, the usage frequency of the device.
Manufacturers of class C and D devices shall update the PSUR at least annually. That PSUR shall be part of the technical documentation as specified in Annexes II and III.
This tool is provided for convenience only and has no legal effect. Only the texts published in the Official Journal of the European Union are authentic. Source: EUR-Lex, CELEX 02017R0746-20250110, © European Union. Reuse is authorised provided the source is acknowledged.